CA-900 Automatic 3-part Differential Hematology Analyzer (CA-900) is an instrument with advanced ARM technology applied in processor which ensures the accuracy and efficiency of the measurement.
CA-900 counts by electrical impendence method. Three kinds of reagents are needed for a complete test. Dilution can be performed automatically. CA-900 can do 3-part differential on WBC, and test 22 parameters like WBC, RBC, HGB, PLT, MCV, etc. in high accuracy. The compact and fully integrated design of CA-900 brings convenience to operation. 8.4 inch TFT color LCD screen allows rapid and accurate patient data entry and monitoring of the results. 35000 results (histograms included) can be stored which is ideal for high volume laboratories. Operation is simple with a WINDOWS-like operation menu system. All regular errors can be detected by the instrument automatically which is helpful to the operator. External printer is optional for print out of test reports.
uitable for second-class hospital, laboratory with moderate workload
Each sample test costs≤RMB1.00
SMC solenoid valve, reliable, rapid, accuracy, stable
Equipped with hydraulic sensors, detect the volume of reagents automatically
Oversized color LCD screen, online service in both Chinese and English
Function of data edition and management, inputting instantly in both Chinese and English
Clog can be solved efficiently by intelligent reverse flushing and burning
Leading technologies in the industry with mature hydraulic design
Operating Principle:Impendence Method &Cyanide-free HGB measure
Test Parameter:3-part differential on WBC,22 test parameter including 3 histograms
Screen:8.4 inch color LCD screen, online service in both Chinese and English
Sample Volume:20ul cap,10ul vein
Throughout:60tests/hour,24hours full time power on with auto standby
Detection System:Double Channels
Language:Chinese/English
I/O Interface:RS232、standard USB、USB-B、PS/2、LAN available
Report Printing:Built-in printer, external printer(optional)
Power Supply:AC220V±10% 50HZ
Reproducibility(CV):WBC CV≤2.0%、RBC CV≤1.5%、MCV CV≤0.5%
PLT CV≤2.5%、 HGB CV≤1.5%
Linearity Range:WBC 0-99.9×109/L、RBC 0-9.9×1012/L、HGB 0-300g/L、
PLT 0-999×109/L、MCV 40-150fL
Carry over: WBC、RBC、HGB、<0.5% PLT≤1.0%
Certification:SFDA、ISO9001、ISO13485、CE、CMC、CMD
Testing Parameter
WBC Leukocyte count
LYM# Lymphocyte count
MIN# Monocytes count
GRA# Granulocyte count
LYM% Lymphocyte percentage
MIN% Monocytes percentage
GRA% Granulocyte percentage
RBC Erythrocyte count
HGB Hemoglobin
HCT Hematocrit
MCV Mean erythrocyte volume
MCH Mean cell hemoglobin
MCHC Mean cell hemoglobin concentration
RDW-CV RBC distribution width (CV)
RDW-SD RBC distribution width (SV)
PLT Platelet count
PCT Plateletcrit
MPV Mean platelet volume
PDW Platelet distribution width
WBC、RBC、PLT 3 histogram